1. Inleiding
The INSTRUKART Keller LEO2 Digital Pressure Gauge is a compact, battery-powered instrument designed for precise pressure measurement in various industrial applications, servicing, and calibration/testing. It features a highly accurate piezoresistive pressure cell, individually adjusted for optimal performance, ensuring a high level of accuracy. This manual provides essential information for the safe and effective use of your LEO2 Digital Pressure Gauge.
2. Veiligheidsinligting
WARNING: Please read this manual carefully before operating the device. Failure to follow instructions may result in damage to the device or personal injury.
- Moenie die toestel blootstel aan uiterste temperature, humiditeit of korrosiewe omgewings nie.
- Ensure the pressure connection is securely fastened before applying pressure.
- Do not exceed the specified pressure ranges.
- Replace batteries only with the recommended type (CR 2430) and ensure correct polarity.
- Do not attempt to disassemble or repair the device yourself. Refer servicing to qualified personnel.
3. Pakketinhoud
Verifieer dat al die items wat hieronder gelys word, in u pakket teenwoordig is:
- 1 x Keller LEO2 Digital Pressure Gauge
- 1 x handleiding
- 1 x G 1/4" Adapter
4. Produkkenmerke
The Keller LEO2 Digital Pressure Gauge offers a range of advanced features for reliable pressure measurement:
- Microprocessor-Controlled: Ensures highly accurate and versatile pressure measurement.
- Digitale aanduiding: Clear display of actual pressure, along with maximum (Max.) and minimum (Min.) pressure values recorded since the last reset.
- Piezoresistive Pressure Transducer: Equipped with an internal EEPROM for precise and stable readings.
- Wide Pressure Ranges: Available in various ranges, including 0 to 100 bar (resolution 1 mbar), -1 to 30 bar (resolution 10 mbar), 0 to 300 bar (resolution 100 mbar), and 0 to 700 bar (resolution 200 mbar).
- Lang batterylewe: Provides up to 1000 hours of continuous operation using 3V CR 2430 batteries.
- Hoë akkuraatheid: Less than 0.1 %FS (Full Scale).
- Robuuste ontwerp: IP 65 protection rating, suitable for various industrial environments.
- Kompakte afmetings: Measures 59 x 95 x 32 mm and weighs 125 g.
- Standard Pressure Connection: Features a 7/16"-20 UNF connection, with a G 1/4" adapter included for versatility.
5. Produk verbyview

Figuur 1: Voorkant view of the LEO2 Digital Pressure Gauge, displaying current and maximum pressure values.

Figuur 2: Kant view illustrating the robust pressure connection at the base of the gauge.

Figuur 3: Voorkant view of the gauge with the G 1/4" adapter securely attached.

Figuur 4: Hoekig view highlighting the clear digital display and user interface buttons.

Figuur 5: Agter view, indicating the location of the battery compartment for CR 2430 batteries.

Figure 6: The LEO2 gauge connected to a flexible hose, demonstrating its adaptability for various setups.

Figure 7: The LEO2 Digital Pressure Gauge neatly stored within its protective carrying case.
6. Opstelling
6.1. Battery Installasie
- Locate the battery compartment cover at the rear of the device (refer to Figure 5).
- Maak die deksel versigtig oop.
- Insert the 3V CR 2430 batteries, ensuring correct polarity as indicated inside the compartment.
- Sluit die deksel van die batterykas veilig.
6.2. Pressure Connection
- The LEO2 features a 7/16"-20 UNF pressure connection. An adapter for G 1/4" is included for broader compatibility.
- Attach the appropriate adapter (if needed) to the gauge's pressure connection.
- Securely fasten the gauge (with or without adapter) to your pressure source. Ensure all connections are tight and leak-free. Use appropriate sealing materials (e.g., PTFE tape) if necessary to prevent leaks.
7. Operasie
7.1. Skakel aan/af
The LEO2 Digital Pressure Gauge powers on automatically when any button is pressed. To conserve battery life, the device will automatically power off after a period of inactivity.
7.2. Display Readings
The gauge features a dual display:
- Topskerm: Shows the current actual pressure reading.
- Onderste skerm: Shows either the maximum (MAX) or minimum (MIN) pressure recorded since the last reset.
7.3. Selecting Display Mode (MAX/MIN)
Use the 'SELECT' button (typically on the left) to toggle between displaying the MAX and MIN pressure values on the bottom screen.
7.4. Resetting MAX/MIN Values
To clear the recorded MAX and MIN pressure values, press and hold the 'ENTER' button (typically on the right) until the values reset.
8. Onderhoud
8.1. Battery vervanging
When the battery indicator appears on the display or the device fails to power on, it is time to replace the batteries. Refer to section 6.1 'Battery Installation' for detailed instructions.
8.2. Skoonmaak
Maak die buitekant van die toestel skoon met 'n sagte, damp cloth. Do not use abrasive cleaners, solvents, or harsh chemicals, as these can damage the casing or display. Ensure no liquid enters the pressure connection or display area.
8.3. Berging
When not in use for extended periods, store the LEO2 Digital Pressure Gauge in its protective carrying case (refer to Figure 7) in a dry, cool place, away from direct sunlight, extreme temperatures, and corrosive substances.
9. Probleemoplossing
If you encounter issues with your LEO2 Digital Pressure Gauge, refer to the following common troubleshooting steps:
- No Display / Device Not Powering On:
- Gaan batterylading na.
- Maak seker dat batterye korrek geïnstalleer is met die regte polariteit.
- Replace batteries with new CR 2430 cells if necessary.
- Onakkurate lesings:
- Verify that the pressure connection is secure and free of leaks.
- Ensure the device is operating within its specified temperature and pressure ranges.
- Check for any physical damage to the sensor or connection.
- Display Errors or Unresponsive Buttons:
- Try resetting the device by temporarily removing and then reinserting the batteries.
- As die probleem voortduur, kontak kliëntediens vir verdere hulp.
10. Spesifikasies
| Kenmerk | Spesifikasie |
|---|---|
| Handelsmerk | INSTRUKART |
| Model | LEO2 |
| Produk afmetings | 3.74 x 2.32 x 1.26 duim (59 x 95 x 32 mm) |
| Item gewig | 125 gram (4.41 onse) |
| Materiaal | Stainless Steel (for pressure connection) |
| Akkuraatheid | < 0.1 %FS |
| Bedryfstemperatuur | -20 to 70 °C / 0 to 50 °C |
| Kragtoevoer | 3V batteries, Type CR 2430 |
| Batterylewe | 1000 uur aaneenlopende werking |
| Beskermingsgradering | IP 65 |
| Drukverbinding | 7/16”-20 UNF (Adapter G 1/4” included) |
| Pressure Ranges (Examphulle) | 0 to 100 bar (Resolution: 1 mbar) -1 to 30 bar (Resolution: 10 mbar) 0 to 300 bar (Resolution: 100 mbar) 0 to 700 bar (Resolution: 200 mbar) |
11. Waarborg en Ondersteuning
For detailed warranty information, technical support, or service inquiries regarding your INSTRUKART Keller LEO2 Digital Pressure Gauge, please contact INSTRUKART customer service. Please retain your purchase receipt as proof of purchase for warranty claims.