PeakTech 5086
PeakTech 5086 Digital LED Lux Meter User Manual
Model: 5086 | Brand: PeakTech
1. Inleiding en oorview
The PeakTech 5086 is a precise 3 3/4 digit (max. 3999) LCD digital LED Lux Meter designed for measuring the light intensity of various light sources, including LED, fluorescent, and traditional incandescent lamps with different light colors. It measures in Lux (lx), Footcandle (FC), or Candela (cd).
Unlike simpler lux meters that are typically calibrated for incandescent bulbs at 2856 K and provide inaccurate results for different color temperatures, the PeakTech 5086 allows for the adjustment of various light source factors. This enables accurate measurements for neon lamps, gloeilamp lamps, red, blue, yellow LEDs, or daylight. The device features 10 freely adjustable correction factors for different light sources, 99 memory locations for measurement values, and useful measurement functions such as minimum, maximum, and average value tracking, as well as data hold. This makes it suitable for professional applications in occupational safety or lighting system evaluation.
Sleutel kenmerke:
- Measures light intensity in Lux (lx), Footcandle (FC), or Candela (cd).
- 10 freely adjustable light source factors for various lamp tipes.
- Suitable for colored light sources, LEDs, and other bulb types.
- Measurement refresh rate: approximately 2.5 times per second.
- External sensor with cosine angle correction.
- Data Hold, Minimum/Maximum/Average (AVG) measurement functions.
- Automatic range selection and zero adjustment.
- 99 memory locations for storing measurement values.
2. Veiligheidsinligting
Lees en verstaan asseblief alle veiligheidsinstruksies voordat u die toestel gebruik. Versuim om hierdie instruksies te volg, kan lei tot besering of skade aan die toestel.
- Moenie die toestel aan uiterste temperature, humiditeit of direkte sonlig blootstel nie.
- Vermy sterk vibrasies of impakte.
- Moenie probeer om die toestel oop te maak of te wysig nie. Verwys alle diens na gekwalifiseerde personeel.
- Maak seker dat die batterye met die korrekte polariteit ingesit is. Verwyder die batterye as die toestel vir 'n lang tydperk nie gebruik gaan word nie.
- Keep the sensor clean and free from obstructions to ensure accurate readings.
3. Pakketinhoud
Verifieer dat al die items wat hieronder gelys word, in u pakket teenwoordig is:
- PeakTech 5086 Digital LED Lux Meter with external sensor
- Drasak
- Batterye (vooraf geïnstalleer of apart ingesluit)
- Gebruikershandleiding
4. Toesteluitleg
The PeakTech 5086 consists of a main unit with an LCD display and control buttons, and an external light sensor connected via a coiled cable. The sensor is equipped with a protective cap.

Figuur 4.1: PeakTech 5086 Digital LED Lux Meter showing the main display unit and the external light sensor with its protective cap removed.
Kontroles en skerm:
- LCD-skerm: Shows measurement values, units (Lux, FC, Cd), battery status, and active functions (e.g., HOLD, MIN, MAX, AVG, Light Source factor).
- Kragknoppie (⏻): Skakel die toestel aan of af.
- LX/FC/CD Button: Toggles between Lux, Footcandle, and Candela measurement units.
- MEM/READ Button: Accesses memory functions to store or recall readings.
- HOLD/L.S. Button: Activates Data Hold function or enters Light Source factor adjustment mode.
- NUL-knoppie: Performs zero adjustment for the sensor.
- op (▲) / Af (▼) Knoppe: Used for navigation and adjusting values in settings modes.
5. Opstelling
5.1 Battery installasie
The PeakTech 5086 typically comes with batteries pre-installed. If not, or if batteries need replacement:
- Vind die batteryklepdeksel aan die agterkant van die hoofeenheid.
- Open the cover by sliding or unscrewing it (refer to the diagram in the full manual if unsure).
- Insert the required batteries (usually AAA or 9V, check the compartment for markings) ensuring correct polarity (+/-).
- Maak die batterykompartement se deksel stewig toe.
5.2 Verbinding van die sensor
The external sensor is permanently attached via a coiled cable. Ensure the cable is not damaged or excessively stretched. The sensor head should be clean and the protective cap removed before taking measurements.
6. Bedryfsinstruksies
6.1 Aan-/afskakel
- Druk die aan / af-knoppie (⏻) to turn the device on. The display will illuminate and show a reading.
- Druk en hou die Power-knoppie (⏻) for a few seconds to turn the device off.
6.2 Unit Selection (Lux/FC/Cd)
- Druk die LX/FC/CD button to cycle through the available measurement units: Lux (lx), Footcandle (FC), and Candela (cd). The selected unit will be displayed on the LCD.
6.3 Nul-aanpassing
- Before taking measurements, especially in low light conditions, it is recommended to perform a zero adjustment.
- Place the protective cap over the sensor to block all light.
- Druk die NUL button. The display should show "0" or a very small value.
- Remove the protective cap before measuring.
6.4 Setting Light Source Factors (L.S.)
The PeakTech 5086 allows for compensation of different light sources. There are 10 adjustable factors (L.S. 0 to L.S. 9).
- Druk en hou die HOLD/L.S. button until "L.S." appears on the display, indicating you are in Light Source factor adjustment mode.
- Gebruik die op (▲) en Af (▼) buttons to select one of the 10 factor memory locations (L.S. 0 to L.S. 9).
- Druk HOLD/L.S. again to enter the adjustment for the selected factor.
- Gebruik die op (▲) en Af (▼) buttons to adjust the factor value. Consult the full manual or light source specifications for appropriate values.
- Druk HOLD/L.S. to confirm the setting and exit the adjustment mode.
6.5 Data Vashou Funksie
- Tydens 'n meting, druk die HOLD/L.S. button (short press) to freeze the current reading on the display. "HOLD" will appear.
- Druk HOLD/L.S. again to release the hold and resume live measurements.
6.6 Minimum, Maximum, and Average (AVG) Measurement
This function allows you to track the lowest, highest, and average light intensity over a measurement period.
- Druk die MEM/LEES button (short press) repeatedly to cycle through MIN, MAX, and AVG modes. The corresponding indicator will appear on the display.
- In MIN mode, the display shows the lowest reading recorded since entering the mode.
- In MAX mode, the display shows the highest reading recorded.
- In AVG mode, the display shows the average reading.
- To exit MIN/MAX/AVG mode, press the MEM/LEES button until no indicator is shown, or turn the device off and on again.
6.7 Memory Functions (Store/Recall)
The device can store up to 99 measurement values.
Storing a Measurement:
- With a live reading on the display, press and hold the MEM/LEES button until "MEM" flashes and a memory location number appears.
- Gebruik die op (▲) en Af (▼) buttons to select an empty memory location (or overwrite an existing one).
- Druk MEM/LEES again to store the value. The display will briefly confirm storage.
Recalling Stored Measurements:
- Druk die MEM/LEES button (short press) repeatedly until "READ" appears on the display.
- Gebruik die op (▲) en Af (▼) buttons to scroll through the stored memory locations.
- To exit recall mode, press MEM/LEES until "READ" disappears, or turn the device off and on.
7. Onderhoud
7.1 Skoonmaak
- Wipe the device and sensor with a soft, dry cloth.
- Moenie skuur skoonmaakmiddels of oplosmiddels gebruik nie.
- Keep the sensor's light-receiving surface clean and free of dust or fingerprints for accurate readings.
7.2 Batteryvervanging
When the low battery indicator appears on the display, replace the batteries as described in Section 5.1. Always use fresh batteries of the specified type.
7.3 Berging
- When not in use, store the device in its carrying bag with the sensor cap in place.
- Berg in 'n koel, droë plek, weg van direkte sonlig en uiterste temperature.
- Remove batteries if storing for an extended period to prevent leakage.
8. Probleemoplossing
| Probleem | Moontlike oorsaak | Oplossing |
|---|---|---|
| Toestel skakel nie aan nie. | Dooie of verkeerd geïnstalleerde batterye. | Kontroleer die polariteit van die batterye en vervang die batterye indien nodig. |
| Onakkurate lesings. | Dirty sensor, incorrect light source factor, or sensor cap still on. | Clean the sensor. Verify the correct light source factor is set. Ensure sensor cap is removed. Perform zero adjustment. |
| Die skerm wys "OL" (Oorlading). | Light intensity exceeds the measurement range. | Move the sensor away from the light source or use in a less intense environment. |
| Cannot store or recall data. | Geheue vol of verkeerde werking. | Ensure you are following the correct steps for storing/recalling. If memory is full, consider overwriting old data or noting down critical values. |
9. Spesifikasies
| Kenmerk | Detail |
|---|---|
| Vertoon | 3 3/4 digit LCD (max. 3999) |
| Meeteenhede | Lux (lx), Footcandle (FC), Candela (cd) |
| Light Source Factors | 10 freely adjustable factors |
| Herlaaitempo | Ongeveer. 2.5 keer per sekonde |
| Soort sensor | External sensor with cosine angle correction |
| Meetstandaard | JIS C 1609:1993 + CNS 5119 Class A |
| Funksies | Data Hold, Min/Max/AVG, Auto Range, Zero Adjustment |
| Geheue | 99 data storage locations |
| Kragbron | Batterye (ingesluit) |
| Afmetings (Pakket) | 21 x 14.3 x 6.7 cm |
| Gewig (Pakket) | 430 gram |
| Modelnommer | Bl GEN 3C 0323 883 |
| ASIN | B017KMZ1H2 |
| EAN/UPC | 04250569403460 |
10. Waarborg en Ondersteuning
PeakTech products are designed for reliability and performance. For specific warranty terms and conditions, please refer to the documentation provided with your purchase or visit the official PeakTech webwebwerf.