1. Inleiding
The Hanna Instruments HI 711 Checker HC is a handheld colorimeter designed for accurate and precise measurement of total chlorine levels. This device offers a simpler and more cost-effective alternative to chemical test kits while providing greater accuracy. It utilizes the EPA-approved DPD method 4500-Cl G, making it suitable for applications requiring this standard. This manual provides essential information for the proper setup, operation, maintenance, and troubleshooting of your HI 711 Checker HC.
2. Veiligheidsinligting
Please read all instructions carefully before using the HI 711 Checker HC. Failure to do so may result in inaccurate readings or damage to the device. Always handle reagents with care and follow all safety precautions provided with the reagent kits. Keep the device and reagents out of reach of children.
3. Produk verbyview
The HI 711 Checker HC is a compact, single-button operation device featuring a large LCD screen for easy readability. It is designed for portability and ease of use in various environments.

Prent 1: The Hanna Instruments HI 711 Checker HC Handheld Colorimeter shown with a sample vial. The device displays a reading of 0.87 ppm for Total Chlorine.
Sleutelkomponente sluit in:
- LCD-skerm: Shows measurement readings and operational prompts.
- Operation Button: Used for powering on/off and initiating measurements.
- Cuvette Holder: Compartment for inserting sample flessies.
- Battery kompartement: Located at the rear for AAA battery installation.
4. Opstelling
4.1 Battery installasie
- Vind die batterykompartement aan die agterkant van die toestel.
- Verwyder die batterydeksel.
- Insert one (1) AAA battery, ensuring correct polarity (+/-).
- Plaas die batterydeksel veilig terug.
4.2 Aanvanklike voorbereiding
Before first use, ensure the cuvettes are clean and free from scratches or debris. Use distilled or deionized water for rinsing.
5. Bedryfsinstruksies
The HI 711 Checker HC operates using a simple, single-button interface. Follow these steps for accurate total chlorine measurements:
- Krag aan: Press the button to turn the Checker HC on. The display will show "---" then "Add C1".
- Zero the Device:
- Fill a clean cuvette with 10 mL of unreacted sample water (the water you wish to test, before adding reagent).
- Cap the cuvette and wipe it thoroughly with a lint-free cloth to remove any fingerprints or smudges.
- Insert the cuvette into the Checker HC, ensuring the cap is fully seated.
- Press the button. The display will show "Add C2" when zeroing is complete.
- Voeg reagens by:
- Remove the cuvette from the Checker HC.
- Add one packet of HI 711-0 Total Chlorine reagent to the sample in die kuvet.
- Cap the cuvette and shake gently for approximately 20 seconds to dissolve the reagent.
- Wipe the cuvette thoroughly with a lint-free cloth.
- Meet:
- Insert the cuvette back into the Checker HC.
- Press the button. The display will show a countdown timer (e.g., "2:00"). This is the reaction time required for the DPD method.
- After the countdown, the Checker HC will display the total chlorine concentration in ppm (parts per million).
- Skakel af: The device features an auto shut-off function to conserve battery life. It will automatically turn off after 10 minutes of inactivity. To manually turn off, press and hold the button until "OFF" is displayed.

Prent 2: The Hanna Instruments HI 711 Checker HC showing a measurement of 0.82 ppm for Total Chlorine on its LCD screen.
6. Onderhoud
6.1 Skoonmaak
- Always keep the cuvettes clean. Rinse them thoroughly with distilled or deionized water after each use.
- Wipe the exterior of the Checker HC with a soft, damp lap. Moenie skuur skoonmaakmiddels of oplosmiddels gebruik nie.
- Ensure the cuvette holder is clean and dry before storage.
6.2 Berging
- Store the Checker HC in a cool, dry place away from direct sunlight and extreme temperatures.
- If storing for extended periods, remove the AAA battery to prevent leakage.
- Store reagents in their original packaging in a cool, dark place.
7. Probleemoplossing
This section addresses common issues you might encounter with your HI 711 Checker HC.
| Probleem | Moontlike oorsaak | Oplossing |
|---|---|---|
| Onakkurate of teenstrydige lesings | Dirty cuvette, expired reagent, improper zeroing, air bubbles in sample | Clean cuvette thoroughly, use fresh reagent, ensure proper zeroing with unreacted sample, remove air bubbles by gently tapping the cuvette. |
| Toestel skakel nie aan nie | Dead battery, incorrect battery installation | Replace AAA battery, check battery polarity. |
| "Err"-boodskap op skerm | Measurement outside range, light interference | Verseker sample is within the device's measurement range. Perform measurement in a stable light environment. Re-zero the device. |
8. Spesifikasies
The following are the technical specifications for the Hanna Instruments HI 711 Checker HC:
- Model: HI-711
- Parameter: Totale chloor
- Metode: Adaptation of the EPA DPD method 4500-Cl G
- Kragbron: 1 x AAA-battery
- Produk afmetings: 5 x 3 x 4 duim
- Gewig: 8.82 onse
- Outo-afskakel: Na 10 minute se onaktiwiteit
9. Waarborginligting
Hanna Instruments products are warranted against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The specific warranty period may vary by region and product. Please refer to the warranty card included with your product or visit the official Hanna Instruments webwebwerf vir gedetailleerde waarborgbepalings en -voorwaardes. Bewaar u bewys van aankoop vir waarborgeise.
10. Kliëntediens
For technical assistance, troubleshooting, or information regarding reagents and accessories, please contact Hanna Instruments customer support. You can find contact details on the official Hanna Instruments website or through your local distributor. When contacting support, please have your product model number (HI-711) and purchase date available.